This topic provides information relating to the CE Declaration of Conformity (Vicon MX T-Series).

Conformity of the Metrological Performance of CLASS 1 Products in accordance with
Annex V and VII Section 5 of the Medical Devices Directive 93/42/EEC as amended by
Directive 2007/47/EC

We, Vicon Motion Systems Limited
Unit 6, Oxford Industrial Park
Mead Road, Yarnton, Oxford, OX5 1QU
UNITED KINGDOM

declare that the VICON MX T-Series motion capture system manufactured by VICON MOTION SYSTEMS LIMITED has been tested prior to shipment and meets the following metrological performance:

  • Resolution of the distance between the centers of two static 14 mm spherical markers located within a volume no less than 4 m x 4 m x 1.5 m to within 1 mm Mean; 1 mm Standard Deviation; sample size no less than 1,000
  • Resolution of a given analog voltage to within +/-20 mV RMS within the following configurations and constraints:
    • No fewer than two cameras of any variant fully viewing static markers
    • Independent of lens and strobe variants fitted to each camera
    • Controlled lighting (no greater than 100 lux) and temperature (17-25° C)
    • Single termination to each analog input
    • Testing using the following Vicon application software: Nexus Version 1.4 or later