  1. Start streaming global angles. (If you're not sure how to do this, see Start Capture.U and choose a capture mode.)

    Tip

    If you want to use real-time feedback to verify the calibration, you can calibrate in Real-Time Insight capture mode and then after calibration, capture global angles in To Sensor mode.

    In all capture modes, when you start a capture, if the sensors need calibrating, a warning message informs you that sensor calibration is required.
    In To Sensor and To Device modes, when you're selecting sensors, or streaming or capturing global angles, IMU sensors that require calibration are displayed with a 'Required calibration' icon at the top right of the sensor . (In Real-Time Insight and AR Visualization modes, the sensor calibration indicator is only displayed when you select the sensors.)  

    Scroll Pagebreak

  2. Wave the sensor in a figure-of-eight for at least 10 seconds, ensuring that it rotates through all three axes during the motion.

    Scroll Table Layout
    orientationdefault
    sortDirectionASC
    repeatTableHeadersdefault
    style
    widths30%, 70%
    sortByColumn1
    tableStylingconfluence
    sortEnabledfalse
    cellHighlightingtrue

    Image Modified
    Image Modified
  3. Put the sensor on a flat surface and spin it through two full rotations around the z axis.

    Tip

    To avoid any unwanted influence on the sensor, avoid placing it in proximity to anything metallic.

  4. Wait at least five seconds until the data has stabilized.
    In Real-Time Insight and AR Visualization modes, you can observe that x angle is 0 and the y angle is 0 (the figure-of-eight provides enough information for the sensor to self-calibrate). However, the z angle may have a non-zero value, depending on how it is rotated to the magnetic north.
  5. Rotate the sensor on the flat surface for at least five seconds  until z is also 0.
    In To Device and To Sensor modes, to let you know that a sensor is calibrated, after 60 seconds, a C (calibration) icon is displayed at the top right of the sensor.
    In Real-Time Insight and AR Visualization modes, check that the graphs for all three axes are now at zero.
  6. After you have calibrated each sensor, you can attach the sensors to your subject and capture data in the required capture mode.

    Note
    If you plan to capture global angles, after the sensors are attached to a subject, allow a few seconds for onboard calibration to initiate before starting the movement that is to be captured.

...