Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

This topic provides information relating to the EU Declaration of Conformity (Vicon MX T-Series).

Conformity of the Metrological Performance of CLASS 1 Products in accordance with
Annex

...

VII, Section 5 of the Medical Devices Directive 93/42/EEC

...

of 14th June 1993.
Aas amended by

...

EU Council Directive 2007/47/EC of 5th September 2007.

We, Vicon Motion Systems Limited
Unit 6, Oxford Industrial Park
Mead Road, Yarnton, Oxford, OX5 1QU
UNITED KINGDOM

declare that the VICON MX T-Series motion capture system SeriesSystem manufactured by VICON MOTION SYSTEMS LIMITED has been tested prior to shipment and meets the following metrological performance:

...

Measurement Criteria

Supporting software Nexus 2.10 or later, Blade 3.0 or later.

Using a minimum of four cameras, resolution of the distance between the centers of two static 14 mm spherical markers located within a volume no less than 4 m x 4 m x 1.5 m to within 1 mm Mean; 1 mm Standard Deviation; sample size no less than 1,000.

Analogue Digital Conversion

Resolution

...

to ± 10 mV mean and ± 10 mV (1 Standard Deviation).

Synchronization

Difference within one video frame.